Gas sampling and conditioning technology 10-1.1-ME Instruction Manual Combustion Air Conditioning Unit Series BA Version BA, BA-P4
10 Gas sampling and conditioning technology 10-1.1-ME Catalytic oxidation Fl3 of the hydrocarbons at a temperature of 500°C on t
11 Gas sampling and conditioning technology 10-1.1-ME 9 DESCRIPTION Figure 2 shows the BA-P4 combustion air conditioning unit. Burner air supp
12 Gas sampling and conditioning technology 10-1.1-ME The catalyst cartridge is installed in a heat-insulated tube furnace. A specia
13 Gas sampling and conditioning technology 10-1.1-ME 10 RECEIPT AND STORAGE The BA... combustion air conditioning unit is a complete, pre-ins
14 Gas sampling and conditioning technology 10-1.1-ME 12 SUPPLY CONNECTIONS 12.1 HOSE CONNECTIONS Connection to the combustion and waste
15 Gas sampling and conditioning technology 10-1.1-ME 12.2 ELECTRICAL CONNECTIONS W A R N I N G ! Incorrect system voltage can damage the
16 Gas sampling and conditioning technology 10-1.1-ME For signalling the group alarm, the contacts 5 and 9 are available in the Sub-D-Plug (see
17 Gas sampling and conditioning technology 10-1.1-ME 14 SWITCH OFF NOTE! The place of installation of the combustion air conditioning uni
18 Gas sampling and conditioning technology 10-1.1-ME 15.1 SWING PISTON COMPRESSOR The swing piston compressor does not need to be removed for
19 Gas sampling and conditioning technology 10-1.1-ME 15.4 CATALYST CARTRIDGE REPLACEMENT Cartridge replacement can take place without the aid
2 Gas sampling and conditioning technology 10-1.1-ME Content
20 Gas sampling and conditioning technology 10-1.1-ME 15.5 ADSORBER CARTRIDGE REPLACEMENT The following steps must be carried out: Relieve
21 Gas sampling and conditioning technology 10-1.1-ME The inner seals can be removed with a pointed tool (e.g. marking tool). When refitting, t
22 Gas sampling and conditioning technology 10-1.1-ME Heating up Heating flashes Ramp function Sensor control 16 USE OF THE TEMPERATURE CO
23 Gas sampling and conditioning technology 10-1.1-ME 16.3 CHANGING THE PARAMETERS OF THE TEMPERATURE REGULATOR The programming of the regula
24 Gas sampling and conditioning technology 10-1.1-ME 16.4 RESET OF THE CONTROLLER A reset out of the temperature lock happens with connected
25 Gas sampling and conditioning technology 10-1.1-ME 17 TROUBLE SHOOTING The table below is intended to indicate possible causes of faults and
26 Gas sampling and conditioning technology 10-1.1-ME 18 SPARE PART LIST Wear, tear and replacement part requirements depend on specific opera
27 Gas sampling and conditioning technology 10-1.1-ME Figure 5 Circuit diagram BA... with controller 703/70304
28 Gas sampling and conditioning technology 10-1.1-ME Figure 6 Circuit diagram BA... with controller TM48
3 Gas sampling and conditioning technology 10-1.1-ME Dear customer, we have made up this operating manual in such a way that a
4 Gas sampling and conditioning technology 10-1.1-ME Head Office M&C TechGroup Germany GmbH Rehhecke 79 40885 Ratingen Germany Tele
5 Gas sampling and conditioning technology 10-1.1-ME 3 SAFETY INSTRUCTIONS Please take care of the following basic safety procedures w
6 Gas sampling and conditioning technology 10-1.1-ME 5 USED TERMS AND SIGNAL INDICATIONS DANGER! This means that death, severe physic
7 Gas sampling and conditioning technology 10-1.1-ME 6 INTRODUCTION The M&C BA... combustion air conditioning unit has been designed speci
8 Gas sampling and conditioning technology 10-1.1-ME 7 FUNCTION Figure 1 shows the function diagram of M&C combustion air conditioning uni
9 Gas sampling and conditioning technology 10-1.1-ME The inlet pressure is maximum 5.8bar and is limited by relief valve YD2 (works setting). Th